MEASUREMENT OF LUBRICANT OIL
MICROVISCOSITY BASED ON RESONANT FREQUENCY SHIFT OF AFM CANTILEVER
Vlad
– Flaviu ZEGREAN
zegrean@fim.usv.ro
Emanuel DIACONESCU
Department of
Applied Mechanics, University “Stefan cel Mare” of Suceava,
ROMANIA
Abstract.
Experimental investigations on microviscosity
of T90 lubricant oil were conducted using an atomic force
microscope. The resonant frequency of the cantilever beams was
measured in air, in pure water and in sample oil. Based on the
resonant frequency shift the viscosity of the lubricant was
calculated using the formula deducted by Papi [1] for uncalibrated
cantilevers. The results obtained are in good agreement with those
of Ionescu [2], measured with a Rheotest2 rheometer, for the same
lubricant.
Keywords:
microviscosity, atomic force microscopy, resonant
frequency shift
References
-
Papi,
M., Arcovito, G., De Spirito, M., Vassalli, M., Tiribilli, B.,
2006, “Fluid Viscosity Determination by Means of Uncalibrated
Atomic Force Microscopy Cantilevers,” Applied Physics Letters,
88, 194102.
-
Ionescu M.,
2004, “A System for Viscosity Measurement in Variable Conditions
of Temperature and Pressure,” MOCM 10, Vol.1, 205 – 208.
-
Binning G., Quate C.F., Gerber Ch.,
1986, “Atomic Force Microscope,” Phys. Rev. Lett. 56, 930
– 933.
-
Sader,
J.E.,
1998, “Frequency Response of Cantilever Beams Immersed in
Viscous Fluids with Applications to the Atomic Force
Microscope,” Journal of Applied Physics, 84, 64 – 76.
-
Chon,
J.W.M., Mulvaney, P., Sader, J.E.,
2000, “Experimental Validation of Theoretical Models for the
Frequency Response of Atomic Force Microscope Cantilever Beams
Immersed in Fluids,” Journal of Applied Physics, 87, 3978
– 3988.
-
Chen,
G.Y., Warmack, R.J., Thundat, T., Allison, D.P., Huang, A.,
1994, “Resonance Response of Scanning Force Microscopy
Cantilevers,” Rev. Sci. Instrum., 65(8), 2532 – 2537.
-
Oden,
P.I., Chen, G.Y., Steele, R.A., Warmack, R.J., Thundat, T.,
1996, “Viscous Drag Measurements Utilizing Microfabricated
Cantilevers,” Applied Physics Letters, 68, 3814 – 3816.
-
Ahmed, N., Nino, D.F., Moy, V.T.,
2001, “Measurement of Solution Viscosity by Atomic Force
Microscopy,” Rev. Sci. Instrum., 72(6), 2731 – 2734.
-
Papi,
M., Maulucci, G., Arcovito, G., Paoletti, P., Vassalli, M., De
Spirito, M.,
2008, “Detection of Microviscosity by Using Uncalibrated Atomic
Force Microscopy Cantilevers,” Applied Physics Letters,
93, 124102.
|